Prepolarized MRI (PMRI)
Prepolarized MRI (PMRI) with pulsed electromagnets has the potential to produce diagnostic quality 0.5- to 1.0-T images with significantly reduced cost, susceptibility artifacts, specific absorption rate, and gradient noise. In PMRI, the main magnetic field cycles between a high field (Bp) to polarize the sample and a homogeneous, low field (B0) for data acquisition. This architecture combines the higher SNR of the polarizing field with the imaging benefits of the lower field. However, PMRI can only achieve high SNR efficiency for volumetric imaging with 3D rapid imaging techniques, such as rapid acquisition with relaxation enhancement (RARE) (FSE, TSE), because slice- interleaved acquisition and longitudinal magnetization storage are both inefficient in PMRI.